SPECIFICATIONS |
|
Operating Voltage |
220 (or 110) Volts +/- 10%, 50 Hz, 1Phase |
|
Power Rating – Heater |
750 Watts |
|
Control System |
Less than 50 Watts |
|
Water Filling / Removal |
Manual. Water to be filled in separate tank |
|
Chamber Size |
2.5 Liters Usable |
|
Control Type |
Micro-computerised control with LCD display |
|
No. Of Fixed Programs |
2 Sterilization Cycles. 134oC 3 mins/18mins. 3 built-in Vacuums |
|
Custom Program |
1 with Password Protection against un-authorized tampering |
|
Programmable Temperature |
120° - 135° C |
|
Time:
|
00 to 59 Min |
|
Vacuums |
1 to 3 times |
|
Special Features |
- User Selectable Pre-Vacuum
- Power Failure Memory - Automatically re-starts selected
Program on Power resumption.
- Password Protection: Prevents unauthorized tampering of set
parameters. Only for Custom program.
- PIN restricted user permission. User Password Generated &
Saved.
- Self Diagnostic Software for Process Fault Recognition
- Manage Users Facility restricted by Admin Password
- Calculates Power Consumed during process
|
|
Alerts Indication |
By Blinking Light |
|
Display |
4 x 20 Liquid Crystal Display (4 Rows x 20 Charatcers) [OPTIONAL:
TOUCH SCREEN] |
|
Indication on Display |
- Program No & Name
- Cycle
- Actual Temp: Deg. C
- Time: Mins
- Help & Process Alerts
- Process Over
- Process Stage in progress
|
|
Power On |
Displays Last Program and Cycle successfully completed |
|
Program Change |
By Up and Down Keys |
|
Reporting |
Report can be generated on Screen for the Last Cycle Run or for Last
5 Cycle Runs. Expandable to 40 cycles
[Optional: Printing] |
|
Parameters Reported |
- Operator Id
- Date & Time of Cycle Start
- Program No. & Name
- Cycle: Wet or Wet & Dry
- Sterilization Temperature
- Sterilization Time
- No. of Pressure Pulse or Vacuums
- Temperature at 1st Pressure Pulse or Vacuum
- Temperature at 2nd Pressure Pulse or Vacuum
- Temp at 3rd Pressure Pulse or Vacuum
- No. of Power Failures
- Process Time Taken
- Power Consumed during the process
- Process Result: Success, Aborted(user stopped) or Unsuccessful
due to process failure with fault indication
|
|
Sterilsable Load |
Solid Hollow or Porous: Wrapped or Unwrapped Hand Pieces |